Recall Z-2207-2023 ยท GXQ Dura Substitute
Medtronic Neurosurgery
Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106; DUREPAIR DURA SUBSTITUTE 4 X 5, 61110; DUREPAIR DURA SUBSTITUTE 1 X 1, 61111; DURA 62100 SUBSTITUTE DUREPAIR 2X2IN NCE, 62100; DURA 62105 SUBSTITUTE DUREPAIR 3X3IN NCE, 62105; DURA 62106 SUBSTITUTE DUREPAIR 1X3IN NCE, 62106; DURA 62110 SUBSTITUTE
| Date | Jul 20, 2023 (initiated Jun 13, 2023) |
| Status | Open, Classified |
| Reason | Dura regeneration matrix may have out-of-specification endotoxin levels due to issues with in-process and finished goods endotoxin testing, which may present clinically with signs and symptoms of an acute inflammatory process, comparable to infection. |
| Root cause | Process change control |
| Action | On 6/13/23 recall notices were mailed to customers and distributors and they were asked to do the following: 1) Return affected product to the recalling firm. 2) Complete and return the Customer Confirmation Form to neuro.quality@medtronic.com 3) This notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Customers with questions can contact Customer Service at 800-874-5797 Option 1. |
| Quantity | 65029 |
| Distribution | Worldwide - US Nationwide distribution including in the states of PR, TX, WA, HI, MI, GA, AL, IL, IN, IA, MO, KY, NY, CA, AZ, PA, UT, ME, MD, MN, FL, AK, NE, VA, OR, WI, NC, LA, SD, TN, MS, DE, NJ, ID, SC, OH, MA, DC, NM, WV, NH, CT, NV, OK, AR, KS, MT, RI, CO, ND, WY, VT and the countries of Barbad |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K161370 | Durepair Dura Regeneration Matrix Medtronic Neurosurgery, Nov 2, 2016 |