ClearanceBrief
Recall Z-2433-2023 ยท DSP System, Balloon, Intra-Aortic And Control

Datascope Corp.

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-00-0800-36, 0998-UC-0800-32, 0998-UC-0800-34, 0998-UC-0800-35, 0998-UC-080

DateAug 30, 2023 (initiated Jul 31, 2023)
StatusOpen, Classified
ReasonUsers reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.
Root causeDevice Design
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/23 was sent to customers. System Over-Temperature User Actions to be taken now: Our records indicate that you may have a Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump (IABP) in your facility. Please examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. When a System Over Temperature alarm is triggered, the device must be powered down andallowed to cool to a safe operating internal temperature before therapy can be re-initiated using the same IABP.Should
Quantity11574 units (5413 US, 6161 OUS)
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181122CARDIOSAVE Intra-Aortic Balloon Pump
Datascope Corp., May 31, 2018