ClearanceBrief
Recall Z-2435-2023 ยท DSP System, Balloon, Intra-Aortic And Control

Datascope Corp.

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65

DateAug 30, 2023 (initiated Jul 31, 2023)
StatusOpen, Classified
ReasonUsers were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kins and abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The HHE also identified hardware issues related to these alarms.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/23 was sent to customers. Gas Loss & Gas Gain Alarms User Actions to be taken now: Our records indicate that you may have a Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump (IABP) affected by this correction in your facility. Please examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. The following actions should be taken should you experience an alarm for Gas Loss in IAB & Gas Gain in IAB : Gas Gain in IAB Circuit " Verify all of the tubing connection
Quantity9175 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181122CARDIOSAVE Intra-Aortic Balloon Pump
Datascope Corp., May 31, 2018