Recall Z-2093-2023 ยท QJR Reagents, 2019-Novel Coronavirus Nucleic Acid

NeuMoDx Molecular Inc

NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and oropharyngeal swabs in transport medium and bronchoalveolar lavage (BAL) specimens from individuals suspected of COVID-19 by their healthcare provider (HCP). REF Number: 300800

DateJul 5, 2023 (initiated May 15, 2023)
StatusOpen, Classified
ReasonPotential for lot number discrepancies between the test strip packaging and the barcode label applied to the test strip that is read by the instrument for LOT 123909 of NeuMoDx SARS-CoV-2 Test Strips (REF 300800). The instrument registers the non-human readable text as lot number 123910 from the barcode label.
Root causeLabeling mix-ups
ActionQIAGEN issued Urgent: Medical Device Correction Letter (PROM-22155-001) on 5/15/23 via email to Lab Director/Manager, Medical Director, Risk Manager,Safety Officer. Letter states reason for recall, health risk and action to take: If you have remaining stock of SARS-CoV-2 Test Strips LOT 123909 REF 300800, do not use it. Please contact QIAGEN Technical Service for a free-of-charge replacement. " Dispose of the product LOT 123909 in accordance with your national and local safety and environmental regulations. " Review this notice with your laboratory/medical director. Important: Forward this inf
Quantity1,086 test strips
DistributionUS Nationwide distribution in the states of FL, MI, NM, NY, TN, TX.
FDA recordOpen on fda.gov