Recall Z-2073-2023 ยท FTM Mesh, Surgical
TEI Biosciences, Inc.
SurgiMend MP; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 10x15 cm, 20x20 cm, 13x25 cm, 16x20 cm, 20x25 cm, 20x30 cm, and 25x40 cm.
| Date | Sep 30, 2024 (initiated May 23, 2023) |
| Status | Open, Classified |
| Reason | Possible out of specification endotoxin test results due to issues with in-process and finished goods endotoxin testing. |
| Root cause | Under Investigation by firm |
| Action | The firm sent recall notifications to customers via email on 5/23/2023 and by mail on 5/24/2023. The recall notification informs customers that Integra is conducting a removal of all Surgimend PRS, Surgimend PRS Meshed, Surgimend, Surgimend MP, Revize, Revize-X, PriMatrix Dermal Repair Scaffold, and PriMatrix AG Antimicrobial Dermal Repair Scaffold devices due to identified issues with in-process and finished goods endotoxin testing that may result in out of specification endotoxin results. Customers are to check their stock for any lot of products listed in the provided Appendix and immediate |
| Quantity | 3,992 units |
| Distribution | Domestic: Nationwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K162965 | SurgiMend MP Collagen Matrix for Soft Tissue Reconstruction Tei Biosciences, Inc., Feb 16, 2017 |