Recall Z-2144-2023 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Biomedical GmbH

Beckman Coulter DxA Automation System, DxA 5000, REF: B87352, IVD, 200240V, 16A, 50/60HZ, RxOnly

DateJul 14, 2023 (initiated Jun 2, 2023)
StatusOpen, Classified
ReasonA software defect in the DxA Automation System may cause tests to be run on patient samples in tube types not compatible with the requested test, leading to erroneous patient results
Root causeSoftware design
ActionOn 06/12/2023, Beckman Coulter issued an Urgent Medical Device Recall Notification to affected consignees. Beckman Coulter asked consignees to take the following actions: 1. Inspect tubes routed to error regions and confirm that the tube type was appropriate per the assay IFU for the test performed. 2. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. 3. So that we are
Quantity146 systems
DistributionUS: AL CA FL IL IN KY MI MN MO NE NJ NV NY OH PA SC TX WI WV OUS: Australia, Belgium, Canada, China, Croatia, Czechia, Egypt, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Kazakhstan, Korea, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Saudi Arabia, Spain, Switzerland, United K
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190298DxA 5000, DxI 800 Access Immunoassay System, Access Ferritin, Access Folate, Access TSH (3rd IS), Access Vitamin B12
Beckman Coulter Biomedical GmbH, Oct 4, 2019