Recall Z-2438-2023 ยท DSP System, Balloon, Intra-Aortic And Control
Datascope Corp.
Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
| Date | Aug 30, 2023 (initiated Jul 31, 2023) |
| Status | Open, Classified |
| Reason | Users were identifying autofill failure conditions on the devices causing pump stops. |
| Root cause | Under Investigation by firm |
| Action | An URGENT MEDICAL DEVICE CORRECTION dated 7/31/23 was sent to customers. Autofill Alarms User Actions to be taken now: Our records indicate that you may have a Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump (IABP) in your facility. Please examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. As a result of the investigation into the complaints, Getinge would like to reinforce information that is detailed in the Cardiosave Instructions for Use to help minimize the frequency and impact of these alarms: As per the Cardiosav |
| Quantity | 9175 units |
| Distribution | Worldwide - US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181122 | CARDIOSAVE Intra-Aortic Balloon Pump Datascope Corp., May 31, 2018 |