Recall Z-2045-2023 ยท IZL System, X-Ray, Mobile

SEDECAL SA

wDR 2.2 Mobile Digital Diagnostic X-Ray System

DateJun 30, 2023 (initiated May 23, 2023)
StatusOpen, Classified
ReasonThere is a software login in issue that may prevent the user from logging in.
Root causeSoftware change control
ActionA recall notification letter was mailed to customers in the United States by the US distributor, Philips, beginning May 23, 2023. Customers are advised to take the following actions: 1) To prevent the system from becoming inoperable, do not reboot or log out of your system until Philips has implemented the software correction. 2) Circulate this notice to all users of the device so that they are aware of the issue. 3) Complete and return the response form to Philips promptly and no later than 30 days from receipt. A Philips representative will contact consignees to schedule a time for a Field
Quantity1061 units
DistributionDistribution throughout United States including Puerto Rico and US Virgin Islands OUS distribution to Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Czech Republic, Dominican Republic, E
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191813MobileDiagnost wDR 2.2
Sedecal., Sa., Aug 2, 2019