Recall Z-2105-2023 ยท HSN Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer

Wright Medical Technology, Inc.

Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile.

DateJul 7, 2023 (initiated Jun 13, 2023)
StatusOpen, Classified
ReasonThe devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.
Root causeProcess control
ActionThe firm issued a Field Safety Correction Action (FSCA) Medical Devices Vigilance System form to the lead competent authority, ANSM-France, via email on 6/13/2023 for their review and approval for the recall to be carried out in their jurisdiction. The FSCA identified the affected product, a description of the issue, and advice on actions to be taken by the distributor and the user. The advice included to check the inventory to locate the product listed and remove and isolate/quarantine it to prevent accidental use. A business reply form was to be returned by email to confirm receipt of th
Quantity106 devices
DistributionThere was no U.S. distribution, including government and military. Foreign distribution was made to Australia, Belgium, Canada, Chile, China, Colombia, France, Germany, Italy, Mexico, Netherlands, Spain, Sweden, Switzerland, Taiwan, and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123954INFINITY TOTAL ANKLE SYSTEM
Wrightmedicaltechnologyinc, Apr 1, 2013