Recall Z-2257-2023 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Megadyne Medical Products, Inc.

MEGADYNE Suction Coagulators, Product Code 004125 (Suction Coagulator, Handswitching, 10 Fr, 6 inch (15.24cm)), Product Code 004225 (Suction Coagulator, Handswitching, 8 Fr, 6 inch (15.24cm), Product Code 004325 (Suction Coagulator, Handswitching, 12 Fr, 6 inch (15.24cm)), and Product Code 0041BN (Suction Coagulator, Handswitching, 10 Fr, 6 inch (15.24cm), Bulk Non Sterile)

DateJul 28, 2023 (initiated Jun 15, 2023)
StatusOpen, Classified
ReasonAn issue with the Suction Coagulator product family may lead to fluid ingress into the handpiece. Fluid ingress can lead to intermittent device activation, non-activation, or self-activation when plugged into the electrosurgery unit (ESU).
Root causeUnder Investigation by firm
ActionAn "URGENT: MEDICAL DEVICE RECALL (REMOVAL)" letter dated 6/15/23 was sent to customers. If you have additional questions regarding this recall or to report any customer complaints, please contact the Ethicon Resource Department at 1-877-ETHICON (1-877-384-4266). The Ethicon Resource Department is open Monday through Friday, 8:00 AM to 5:00 PM ET. ACTION REQUIRED 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). If you have product subject to this recall, please maintain a copy of this notice with the quarant
Quantity241,457 units
DistributionWorldwide Distributions: US (Nationwide) and OUS (Foreign) in countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, NEW ZEALAND, NORWAY, RUSSIA, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, U
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072559SUCTION COAGULATOR
Megadyne Medical Products, Inc., Oct 5, 2007