Recall Z-2033-2023 ยท DPZ Oximeter, Ear
GE Healthcare Finland Oy
TruSignal Integrated Ear Sensor with Datex Connector, REF TS-E4-N; Oximeter
| Date | Jul 7, 2023 (initiated May 19, 2023) |
| Status | Open, Classified |
| Reason | There is a potential reduction of energy reaching patient during defibrillation, potential contact with unintended voltage, or inaccurate measurement. |
| Root cause | Device Design |
| Action | GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice on 05/19/2023. The notice explained the safety issues associated with the us of the device and requested the actions be taken: Actions to be taken by Customer/User for Safety Issue # 1 and #2: 1. Use an alternate method for SpO2 monitoring such as TruSignal Sensors not impacted by this field action, or an alternate SpO2 device, if possible 2. If alternate methods are not possible, the affected TruSignal SpO2 Sensors can be used for monitoring if they have not been saturated with fluids 3. If defibrillation is necessary, when |
| Quantity | 708 devices |
| Distribution | Worldwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K021955 | TRUTRAK PLUS ENHANCEMENTS TO THE DATEX-OHMEDA 3800 SERIES & 3900 SERIES OXIMETERS & ACCESSORIES Datex-Ohmeda, Inc., Oct 8, 2002 |