Recall Z-2101-2023 ยท FMZ Incubator, Neonatal

DATEX--OHMEDA, INC.

Giraffe OmniBed, with the following Model Numbers: 6650-0004-901, 6650-0001-901, 6650-0003-901, 6650-0002-901, 6650-0094-901, 6650-0001-9901, 6650-0029-901, 6650-0080-901, 6650-0032-901, 6650-0077-901, 6650-001-901, 6650-0006-901, 6650-0129-901, 6650-0063-902, 6650-0130-901, 6650-0082-902, 6650-0061-902, 6650-0131-902, 6650-0034-902, 6650-0104-902, 6650-0066-902, 6650-0042-902, 6650-0143-901, 6650

DateJul 6, 2023 (initiated May 19, 2023)
StatusOpen, Classified
ReasonGE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
Root causeNonconforming Material/Component
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/19/2023 by mail. The notice explained the problem, risk, and requeested the following: Actions to be taken by Customer/User: Inspect the device bedside panels; if the device has the correct latches, continue to use the device. If the device does not have the correct latches, quarantine the device. Inspect any field replaceable bedside panel stock and quarantine for the affected product. GE Healthcare will replace affected latches.
Quantity3,053 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020543OHMEDA MEDICAL GIRAFFE OMNIBED
Ohmeda Medical, May 9, 2002
K101788GIRAFFE OMNIBED
Ohmeda Medical, Jul 21, 2010