Recall Z-2100-2023 ยท FMZ Incubator, Neonatal
DATEX--OHMEDA, INC.
Giraffe OmniBed Carestation CS1, Model Numbers: a. 2082844-001-XXX, b. 2074920-001-XXXXXX, c. M1050205, d. M1135822, e. M1145475, f. M1145476, g. M1235957, h. M1149051, i. M1146098; Labeled as the following: GIRAFFE OMNIBED CARESTATION CS1, GIRAFFE OMNIBED CARESTATION, GIRAFFE OMNIBED, GIRAFFE OMNIBED CS1, OMNI G 115/230V SERVO2 CARESTATION, ASSY-LAU, OMNIBED CHINA EBASE S HUMID, Man
| Date | Jul 6, 2023 (initiated May 19, 2023) |
| Status | Open, Classified |
| Reason | GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices |
| Root cause | Nonconforming Material/Component |
| Action | GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/19/2023 by mail. The notice explained the problem, risk, and requeested the following: Actions to be taken by Customer/User: Inspect the device bedside panels; if the device has the correct latches, continue to use the device. If the device does not have the correct latches, quarantine the device. Inspect any field replaceable bedside panel stock and quarantine for the affected product. GE Healthcare will replace affected latches. |
| Quantity | 4,540 units |
| Distribution | Worldwide distribution. |
| FDA record | Open on fda.gov |