Recall Z-2024-2023 ยท OZN C. Difficile Toxin Gene Amplification Assay

Luminex Corporation

VERIGENE Clostridium difficile Nucleic Acid Test, REF 30-002-022

DateJun 28, 2023 (initiated Jun 5, 2023)
StatusOpen, Classified
ReasonIt is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result due to hydrophobic characteristics in specific FLOQSwab lots.
Root causeOther
ActionLuminex issued an Urgent Field Safety Notice - Medical Device Recall notice on 06/05/2023 by email. The notice explained the problem with the stool prep kit, the risk, and requested isolation/quarantine, cease of use, and disposal of the affected products.
Quantity114 units
DistributionWorldwide and US Nationwide Distribution