ClearanceBrief
Recall Z-1997-2023 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Megadyne Medical Products, Inc.

MEGADYNE MEGA SOFT Pediatric Patient Return Electrode. used during electrosurgery

DateJun 30, 2023 (initiated Jun 1, 2023)
StatusOpen, Classified
ReasonFirm has received reports of patient burns in surgical procedures where device was used.
Root causeUnder Investigation by firm
ActionThe firm distributed a recall notification by mail starting 06/01/2023. Customers are to provide the notice to all users of MEGADYNE MEGA SOFT and MEGA 2000 Reusable Patient Return Electrodes that participate in the cleaning, setup, and operation of affected devices. Megadyne has been made aware of 63 reports of serious patient burns since April 2018. The firm has attributed the incidence of burns to some instances that the Mega Soft Pad Instructions For Use were not being properly followed. Devices must be thoroughly rinsed after cleaning to ensure any cleaning residue is removed. Customer
Quantity21,100 units
DistributionDistributed US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080741MEGA SOFT REUSABLE PATIENT RETURN ELECTRODE
Megadyne Medical Products, Inc., Dec 16, 2008