Recall Z-2270-2023 ยท IYB Tube Mount, X-Ray, Diagnostic

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION

Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safire, Versa and G4 units, FluoroSpeed 300 and X1 units

DateJul 31, 2023 (initiated Apr 10, 2023)
StatusOpen, Classified
ReasonEight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to eliminate the risk of the device falling.
Root causeDevice Design
ActionOn 4/10/23, recall notices were emailed to customers who were asked to do the following: 1)The Tube Column Assembly is secured by eight fixing bolts located at the base of the assembly. If while using the system, you observe loose, or missing bolts then please discontinue use of the system and call your authorized Service Provider. 2) Schedule a time with their Distributer to have their system updated. 3) Post the notice within their facility so that all users are aware of how to avoid this issue until the corrective actions has been completed. 4) Complete and return the Customer Acknowledgeme
Quantity1836
DistributionU.S. Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.