Recall Z-1643-2023 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Maquet Medical Systems USA
QUADROX-iR, used during cardiopulmonary bypass Model Item No. Description BEQ-HMO 51100-USA 701067936 Small Adult with filter, with BIOLINE Coating BEQ-HMO 71100-USA 701067880 Adult with filter, with BIOLINE Coating HMO 50100-USA 701067934 Small Adult without filter, with SOFTLINE Coating HMO 51100-USA 70106
| Date | May 26, 2023 (initiated May 18, 2023) |
| Status | Open, Classified |
| Reason | 1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia. |
| Root cause | Pending |
| Action | On May 18, 2023, the firm notified customers of the recall via an Urgent Medical Device Removal letter. Actions to be taken by the customer: If an affected device is already in use, please use according to normal practices. Monitor the patient for signs and symptoms of inflammation, infection, sepsis, and ischemia. If detected, treat according to clinical protocols. Please remove any affected product in inventory from areas of use. Unopened and unexpired affected product is eligible for credit upon return. Product Return: Please contact Getinge Customer Service at 1-888-943-8872 option 2 b |
| Quantity | 0 (US) |
| Distribution | Domestic distribution nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K102726 | CARDIOHELP SYSTEM Maquet Cardiopulmonary, AG, Apr 4, 2011 |