ClearanceBrief
Recall Z-2264-2023 ยท JAK System, X-Ray, Tomography, Computed

Olympus Corporation of the Americas

Always-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8mm OD, Serrated Cup-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-0372

DateJul 28, 2023 (initiated May 15, 2023)
StatusOpen, Classified
ReasonElectromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
Root causeComponent design/selection
ActionOlympus issued "URGENT - Medical Device Field Corrective Action" on 5/15/23. Letter states reason for recall, health risk and action to take: Olympus implemented a new internal component and has observed a reduction in complaints for the above-mentioned issue. However, to minimize the duration of potential procedure delay, Olympus is revising the Instruction for Use to add the below precaution statement: It is required to have additional Always-On Tip Tracked instruments available for all procedures as tracking may be lost when: Connecting the instrument to the SPiN Thoracic Navigation Syste
Quantity3692.2 units
DistributionWorldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: CN, DE, HK, IT, KR, SG, TH & TW.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122106SPIN DRIVE ALWAYS-ON TIP TRACKED STEERABLE CATHETER ALWAYS-ON TIP TRACKED SHEATH ALWAYS-ON TIP TRACKED ASPIRATION NEEDLE
Veran Medical Technologies, Inc., Dec 21, 2012
K170023SPiN Thoracic Navigation System
Veran Medical Technologies, Inc., May 15, 2017