ClearanceBrief
Recall Z-2058-2023 ยท IOL Treadmill, Powered

Full Vision Inc

TRACKMASTER, TREADMILL GE T2100-ST2 220V CHINA, MODEL NUMBER 317-07927GE CHINA

DateJun 30, 2023 (initiated May 12, 2023)
StatusOpen, Classified
ReasonTreadmill malfunction: drive PCB can misread the signals and stop or reverse and then accelerate forward.
Root causeOther
ActionTrackmaster issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/12/2023 by letter. The notice explained the problem and the risk during use, and requested the following: Immediate Actions to be taken by Customer/User: These steps are the recommendation of the Manufacturer to prevent injury and allow use of the device. Step 1: Instruct patient to straddle the belt and hold on to the handrails during initialization of the treadmill before starting the test Step 2: Send start/run command to the treadmill with patient still straddling the belt, wait 30 seconds to
Quantity95 units
DistributionUSA (Nationwide), Australia, Bahrain, Canada, Chile, Colombia, Costa Rica, Dominican Republic, FINLAND, Guatemala, HONG KONG, India, Indonesia, Israel, Mexico, NEW ZEALAND, Nicaragua, Panama, PARAQUAY, Peru, Philippines, Saudi Arabia, Singapore, SOUTH KOREA, Thailand, Ukraine, United Arab Emirates,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.