ClearanceBrief
Recall Z-1808-2023 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

ICD-VR VISIA MRI AF S US/OUS DF4, Model Number DVFC3D4; Implantable Cardioverter Defibrillators

DateJun 16, 2023 (initiated May 10, 2023)
StatusOpen, Classified
ReasonThere is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Root causeUnder Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who ha
Quantity16244 units
DistributionUS Nationwide - Worldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.