Recall Z-1758-2023 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

CRTD DTMA2D4 CLARIA MRI OUS DF4, Model Number DTMA2D4; Implantable Cardioverter Defibrillators

DateJun 16, 2023 (initiated May 10, 2023)
StatusOpen, Classified
ReasonThere is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Root causeUnder Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who ha
Quantity4593 units
DistributionUS Nationwide - Worldwide Distribution