Recall Z-1708-2023 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
ICD-DR DDBB2D4 EVERA XT IS1/DF4 INTL, Model Number DDBB2D4; Implantable Cardioverter Defibrillators
| Date | Jun 16, 2023 (initiated May 10, 2023) |
| Status | Open, Classified |
| Reason | There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds. |
| Root cause | Under Investigation by firm |
| Action | Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who ha |
| Quantity | 727 units |
| Distribution | US Nationwide - Worldwide Distribution |
| FDA record | Open on fda.gov |