Recall Z-1993-2023 ยท DSQ Ventricular (Assist) Bypass
Heartware, Inc.
HeartWare Ventricular Assist Device System. a) Monitor 1521US Explore Tech, b) Monitor 1521GB Explore Tech, c) Monitor 1521IL Explore Tech, d) Monitor 1521DE Explore Tech
| Date | Jun 22, 2023 (initiated May 16, 2023) |
| Status | Open, Classified |
| Reason | Medtronic has identified that the logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and MON5xxxxxx for US), are unable to be processed by the Autologs web portal. |
| Root cause | Software change control |
| Action | Medtronic Mechanical Circulatory Support issued an URGENT MEDICAL DEVICE COMMUNICATION notice to it consignees on 05/16/2023 by email. The notice explained that logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and MON5xxxxxx for US), are unable to be processed by the Autologs web portal. Medtronic is working on an update to the Autologs web portal to resolve the issue. Until the update is complete, logfiles can be sent to dl.mcsclinicalengineering@medtronic.com for report generation. Please note in the email that you have received a logf |
| Quantity | 19 units |
| Distribution | Worldwide - US Nationwide distribution in the states of TX, MO, VA, and the countries of Bahrain, UK. |
| FDA record | Open on fda.gov |