Recall Z-1918-2023 ยท PNG Peritoneal, Drainage Catheter For Refractory Ascites, Long-Term Indwelling

Merit Medical Systems, Inc.

Aspira Drainage Catheter Insertion Tray 15.5F, REF: ASP-11-5, STERILE EO, Rx Only *EU Distribution only

DateJun 14, 2023 (initiated Apr 22, 2023)
StatusOpen, Classified
ReasonThere is a potential that valve assemblies will not open. preventing fluid from draining.
Root causeProcess control
ActionBeginning on April 22, 2023, Merit Medical issued an "Urgent Medical Device Recall Notice" to affected consignees via FedEx and UPS. On May 31, 2023, Merit Medical expanded the scope and notified additional consignees. Merit Medical asked consignees to take the following actions: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has bee
Quantity7 units
DistributionWorldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico and the countries of France, United Ki
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110396ASPIRA PERITONEAL DRAINAGE SYSTEM
C.R. Bard, Inc., May 6, 2011
K212675Aspira Peritoneal Drainage System
Merit Medical Systems, Inc., Dec 14, 2021