Recall Z-2152-2023 · DSP System, Balloon, Intra-Aortic And Control
Datascope Corp.
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85
| Date | Jul 20, 2023 (initiated Jun 5, 2023) |
| Status | Open, Classified |
| Reason | An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy. |
| Root cause | Component design/selection |
| Action | On June 5, 2023, the firm notified customers of the recall via "URGENT MEDICAL DEVICE CORRECTION" letters. Letters covered two issues: Issue 1 - Failure to charge battery (Covered by RES 92287) Issue 2 - Unexpected shutdown (Covered by this RES event) User Actions: Issue 2: Unexpected Shutdown due to failure of Tantalum Capacitors. Should loss of power occur, therapy will be interrupted and the console will need to be exchanged with another (if available) to resume therapy. If your device remains inoperable, please contact your service representative. It is recommended for users to ha |
| Quantity | 9174 (4586 US; 4588 OUS) in total |
| Distribution | Worldwide Distribution- US (Nationwide) and OUS (foreign) countries of: ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, AZERBAIJAN , BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿t |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181122 | CARDIOSAVE Intra-Aortic Balloon Pump Datascope Corp., May 31, 2018 |