Recall Z-1970-2023 ยท KXG Applicator, Absorbent Tipped, Sterile

LumiraDx

SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB

DateJun 15, 2023 (initiated Apr 25, 2023)
StatusOpen, Classified
ReasonFails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab
Root causeProcess control
ActionLumiraDx issued on 25 April 2023 via email electronic Cover Letter and Urgent Product Recall Notice with an attestation form to document product destruction at their facility. Letter states reason for recall, health risk and action to take: Immediately stop using the affected lot of swabs " Remove from your stock any remaining affected kits of swabs. " Destroy all stock of affected swabs " Indicate the number of packs that were destroyed, and sign the attestation form at the end of this notice verifying destruction of the affected swabs in your inventory. " Report any issues or adverse events
Quantity1003 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.