ClearanceBrief
Recall Z-2115-2023 ยท OZD Temporary Non-Roller Type Left Heart Support Blood Pump

Abiomed, Inc.

Impella CP intravascular micro axial blood pump, Product Number 0048-0032

DateJul 14, 2023 (initiated Jun 14, 2023)
StatusOpen, Classified
ReasonThere is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.
Root causeUnder Investigation by firm
ActionAn URGENT: MEDICAL DEVICE CORRECTION (NOTIFICATION) dated 6/14/23 was sent to customers. RECOMMENDATIONS: Clinicians are cautioned to position the Impella system carefully in patients with TAVR, and to be aware of this potential interaction. In this situation, clinicians should avoid repositioning while the device is spinning and should turn the device to P0 during repositioning or any movement that could bring the outlet windows into proximity to the valve stent structures. If there is low flow observed in a patient implanted with a TAVR while on Impella heart pump support, you should consi
Quantity9252 units
DistributionWorldwide distribution - US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.