ClearanceBrief
Recall Z-2007-2023 ยท FNL Bed, Ac-Powered Adjustable Hospital

Baxter Healthcare Corporation

NaviCare Patient Safety V3.9.200 to 3.9.600, used with the Centrella Bed

DateJun 26, 2023 (initiated May 19, 2023)
StatusOpen, Classified
ReasonBaxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off.
Root causeDevice Design
ActionBaxter issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/19/2023 by letter (USPS). The notice explained the issue, hazard, and requested the following: Baxter is asking that you take the following actions: 1. Share this letter with your IT department (or other relevant personnel) to determine if you have the impacted software version and contact Baxter Technical Support at 1-800-445-3720, available twenty-four hours a day, seven days a week, to schedule a software upgrade which addresses the identified issue. 2. You may continue to use your Voalte Patient Safety Soft
Quantity7 software versions
DistributionUS Nationwide distribution in the states of OR, SC, VA, TN, GA, NC, CO, NC, FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.