Recall Z-2149-2023 · DSP System, Balloon, Intra-Aortic And Control

Datascope Corp.

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53* 0998-00-0800-55* *Not US distribution

DateJul 19, 2023 (initiated Jun 5, 2023)
StatusOpen, Classified
ReasonIABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the issue prior to interruption of therapy.
Root causeDevice Design
ActionOn June 5, 2023, the firm notified customers of the recall via "URGENT MEDICAL DEVICE CORRECTION" letters. Letters covered two issues: Issue 1 - Failure to charge battery (Covered by this RES event) Issue 2 - Unexpected shutdown (Covered by RES 92335) User Actions: Issue 1: IABP will not charge battery due to failure of the Power Management Board Charge Path Circuitry To prevent electrical surges that may impair the Cardiosave s ability to charge batteries, do not remove the battery from a Cardiosave when the battery level is at 80% or higher and actively charging (i.e successfully connec
Quantity9174 (4586 US; 4588 OUS)
DistributionWorldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: ALBANIA ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181122CARDIOSAVE Intra-Aortic Balloon Pump
Datascope Corp., May 31, 2018