Recall Z-1671-2023 · HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer

Smith & Nephew, Inc.

Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6); Unicondylar knee prothesis

DateMay 31, 2023 (initiated May 2, 2023)
StatusOpen, Classified
ReasonRecent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
Root causeUnder Investigation by firm
ActionSmith & Nephew issued an Urgent Medical Device Recall Notice to its consignees on 05/02/2023 via email and letter. The notice explained the problem and the risk and requested the following: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customer s account number
Quantity1596 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190439Engage™ Partial Knee System
Engage Uni, LLC, Nov 21, 2019