Recall Z-1633-2023 ยท FQZ Thermometer, Clinical Color Change

BearCare, Inc.

Walnut Wearable Smart Thermometer, intended for continuous chest temperature monitoring of children ages 0-6 years in non-emergency medical situations, Model Number WT20

DateJun 8, 2023 (initiated May 10, 2023)
StatusTerminated
ReasonThere have been reports of skin burns to children wearing the device.
Root causeComponent design/selection
ActionAn URGENT: MEDICAL DEVICE RECALL notification letter dated 5/10/23 was sent to customers. Actions to take: BearCare is asking consumers to immediately discontinue use of all Walnut Thermometers employing a rechargeable battery and return the product to BearCare for a refund. BearCare s records indicate approximately nine hundred seventy-nine (979) units are in the possession of consumers. Any consumer noting burns or irritation on their child s skin should seek immediate medical care. Additionally, BearCare is asking consumers to: (1) confirm receipt of the recall notice via email to recall@w
Quantity1818 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.