Recall Z-1604-2023 ยท DZE Implant, Endosseous, Root-Form

Hiossen Inc.

ET III Nano-HA fixture System ET NH Dental Implant Model # ET3R5010B -Intended for Dental Implant Catalog #: AET3R5010B

DateMay 18, 2023 (initiated Apr 6, 2023)
StatusOpen, Classified
ReasonMis-labeled length of the dental implant on the Recall Implant's package. Dental implant size D5.0 x L8.5mm, labeled as D5.0 x L10mm
Root causeLabeling Change Control
ActionHoissen Inc. initial contact verbally through sales rep. on April 7, 2023 and by "URGENT: MEDICAL DEVICE RECALL" Letter via Mail & email on 04/19/2023. Letter states reason for recall, health risk and action to take: 1. Do not use the Recall Implants. 2. Quarantine (separate) the Recall Implants in a different location, away from normal stock room. 3. Immediately fill out the Acknowledgement and Receipt Form* (Fax: 267-795-7031 or Email: recall@hiossen.com) *Response is required regardless if the Recall Implants have already been return or implanted. 4. Returned the Recall Implant(s) as Unusa
Quantity287 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151626ET III Bio-SA Fixture System
Hiossen, Inc., Jul 21, 2016