Recall Z-1628-2023 ยท PDJ Hemoglobin A1c Test System

Ortho-Clinical Diagnostics, Inc.

VITROS Chemistry Products HbA1c Reagent Kit- IVD for the quantitative determination of percent glycated hemoglobin A1c (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human whole blood. Product Code: 6842905

DateMay 24, 2023 (initiated Apr 17, 2023)
StatusOpen, Classified
ReasonPotential for both unexpected positive and negative HbA1c results, may result in misdiagnosis
Root causeUnder Investigation by firm
ActionOrtho Clinical Diagnostics (Ortho) issued to Dealers and End-Users Urgent Product Correction Notification Potential Bias Observed in VITROS Chemistry Products HbA1c Reagent Kit on 4/17/23. Letter states reason for recall, health risk and action to take: In an abundance of caution, the required actions are applicable to all lots within expiry ( as well as future lots until your laboratory receives an additional notification from Ortho. However, while data is limited, a review of the release testing indicated no other lots were impacted. Reagent Pack Used for Calibration o If your system current
Quantity23751
DistributionUS Nationwide- Worldwide Distribution. Foreign: Australia Belgium Brazil Bermuda HM12 Canada L3R 4G5 Chile China Colombia Denmark France France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden The Netherlands United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142595VITROS Chemistry Products HbA 1c Reagent Kit, VITROS Calibrator Kit 31, VITROS Chemistry Products %A1c Performance Verif