Recall Z-2005-2023 ยท ITI Wheelchair, Powered
Permobil
PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application
| Date | Jun 26, 2023 (initiated May 10, 2023) |
| Status | Open, Classified |
| Reason | When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures. |
| Root cause | Software design |
| Action | Permobil began notifying customers via in-person and through letters on 05/10/2023. Customers were instructed to examine their device and update their software to SmartDrive MX2+ Version 1.1.00, to acknowledge the update was conducted and the notification letter was understood through the field action portal, notify customers if the affected devices were further distributed. They were also instructed to discontinue using the affected device while waiting for the correction. |
| Quantity | 6,196 units |
| Distribution | Worldwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K151199 | SmartDrive Wheelchair Power Assist Max Mobility, LLC, Jul 30, 2015 |