Recall Z-1668-2023 ยท OWB Interventional Fluoroscopic X-Ray System

NRT X-RAY A/S

Intelli-C, Right side suspended Tabletop, REF: 03400000

DateMay 31, 2023 (initiated Apr 27, 2023)
StatusOpen, Classified
ReasonThere is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
Root causeDevice Design
ActionBetween the dates of April 27, 2023 - May 1, 2023, NRT X-Ray A/S issued a "urgent: Medical Device Correction" via E-Mail to affected consignees. Per the customer communication, NRT ask consignees to take the following actions: Step 1 Short term action The first step of the recall is a corrective action which requires the existing fixation bracket components to be removed. Service Providers of the Intelli-C systems should follow the instructions in page 4 through 7 of the communication and report back to NRT using the Customer Response Card on page 8. The process is coordinated through US d
Quantity32 units
DistributionUS Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173631Intelli-C
Nrt X-Ray A/S, Jul 11, 2018