Recall Z-1480-2023 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Luminos dRF Max (VE10, VF10, VF11)

DateMay 23, 2023 (initiated Nov 23, 2022)
StatusOpen, Classified
ReasonUnder certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.
Root causeRadiation Control for Health and Safety Act
ActionSiemens Medical Solutions USA, Inc., will provide software updates to the affected systems with software versions VF10 and VF11 with software via Update Instructions (UI) XP007/23/S, XP004/23/S, XP003/23/S, and XP001/23/S
Quantity704 units in the United States (1945 units worldwide)
DistributionUS Nationwide - Worldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173639Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom Rax
Siemens Medi Cal Solutions, Inc., Apr 2, 2018