ClearanceBrief
Recall Z-2260-2023 ยท JAK System, X-Ray, Tomography, Computed

Philips North America

Incisive CT Power (China)-Computed Tomography X-Ray System Model: 728148

DateJul 28, 2023 (initiated Apr 25, 2023)
StatusOpen, Classified
ReasonMetal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
Root causeUnder Investigation by firm
ActionPhiips issued Urgent Medical Device Correction Letters mailed to US consignees via Certified United States Postal Service (USPS) with delivery confirmation on 4/25/23 International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Please continue to use your system in accordance with its intended use. If you need to remain in the room for the duration of a scan, Philips recommends using protective eyewear. " Circulate this Urgent Medical D
Quantity53 units OUS
DistributionNationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Mace
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180015Philips Incisive CT
Philips Healthcare (Suzhou) Co., Ltd., Mar 20, 2018