ClearanceBrief
Recall Z-1972-2023 ยท NQO Prosthesis, Spinous Process Spacer/Plate

Boston Scientific Neuromodulation Corporation

Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual

DateJun 15, 2023 (initiated May 12, 2023)
StatusOpen, Classified
ReasonPending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.
Root causeLabeling design
ActionOn 05/12/23, correction notices were mailed to physicians/surgeons, hospitals, healthcare professionals who were advised the following: IFU advises against forced deployment, as it could result in device breakage. Correction notice advises to review of the IFUs/Surgical Technique Manual. Avoid application of excessive force, use only two or three fingers when performing final tightening during device deployment). Inspect Driver instrument prior to/following use; if Driver tip break is noted: Retrieve metal fragment(s), consider performing imaging. Missing/bent teeth prevents Driver engage
Quantity21,533
DistributionUS Nationwide Distribution: IN, NY, IL, PA, FL, SC, TX, MA, VA, AL, WI, OK, MO, MI, OH, WV, NC, UT, NH, CA, NJ, TN, NV, AZ, NE, IA, CT, ME, LA, KY, MS, MD, DC, GA, CO, KS, ID, MN, AR, HI, NM, WA, SD, OR, DE, WY, AK, and ND
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.