Recall Z-1626-2023 ยท QEP Nucleic Acid Detection System For Non-Viral Microorganism(S) Causing Sexually Transmitted Infections

Binx Health, Inc.

binx io CT/NG Assay- Single-use Cartridge for the qualitative detection of Chlamydia trachomatis and Neisseria gonorrhoeae Ref: 1.002.101

DateMay 23, 2023 (initiated Mar 29, 2023)
StatusOpen, Classified
ReasonMay not meet the declared performance characteristics for the detection of Chlamydia trachomatis (CT) within the labeled expiry period, potential to produce either false negative CT or positive CT results
Root causeUnder Investigation by firm
ActionBinx issued Urgent Medical Device Recall Letter on 29 March 2023 via email. On 31 March 2023 a hard copy of the written Recall Notification and Response Form was sent via certified mail through the United States Postal Service. Letter states reason for recall, health risk and action to take: Discontinue use of the binx io CT/NG Assay lot listed above, and discard your inventory of this lot. " At the Clinic Director s discretion results may be reviewed. RESOLUTION: " binx health has removed this binx io CT/NG Assay lot from our inventory. We will provide alternate lots for all current and futu
Quantity50 single-use cartridges
DistributionIn the US state of PA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200533binx io CT/NG Assay and binx io CT/NG System
Binx Health, Inc., Apr 27, 2020