Recall Z-1594-2023 ยท QKP Coronavirus Antigen Detection Test System.

SD Biosensor, Inc.

Pilot COVID-19 At-Home Test, Reference Number 9901-NCOV-10G, Catalog Number 99COV200L-EN02, UPC Code: 887473000207 (EAN-13: 08800111707728)

DateMay 23, 2023 (initiated Mar 31, 2023)
StatusOpen, Classified
ReasonThe liquid buffer component in the affected test kit lots was determined to have bacterial contamination. User may notice a slight odor when the buffer is brought close to the nose.
Root causeMaterial/Component Contamination
ActionPress release issued 5/4/23 by SD BioSensor: https://www.prnewswire.com/news-releases/recall-notice---sd-biosensor-inc-requests-discontinuation-of-use-and-disposal-of-specific-pilot-covid-19-at-home-tests-in-the-united-states-due-to-microbial-contamination-in-the-liquid-buffer-solution-301816494.html A notification letter titled "Subject: Statement about the patient risk assessment for the Extraction buffer of Pilot" COVID-19 At-Home Test" dated 3/31/23 was sent to Roche. A Roche URGENT MEDICAL DEVICE REMOVAL notification letter dated 5/4/23 was sent to customers. . Assess your current inv
Quantity2,712,767 kits
DistributionUS Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.