Recall Z-1617-2023 ยท NEP System/Device, Pharmacy Compounding

OmniCell, Inc.

Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373

DateMay 22, 2023 (initiated Mar 27, 2023)
StatusOpen, Classified
ReasonIVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling
Root causeUnder Investigation by firm
ActionOmincell notified Customers via phone on 3/27/2023 and by Urgent Field Action: Mandatory Correction Of Product Labeling Ivx Station (Limited Release) Letter on 3/31/2023 via email. Letter states reason for recall, health risk and action to take: We will analyze specific devices at your Limited Release sites. Upon completion of our analysis updated labeling will be provided and will include the revised dose accuracy range.
Quantity6 units
DistributionAL, NY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.