Recall Z-1469-2023 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

ARTIS one with a hybrid cable- An angiography system developed for diagnostic imaging and interventional procedures. Material Number: 10848600

DateApr 27, 2023 (initiated Mar 21, 2023)
StatusOpen, Classified
ReasonExamination room monitor connected with the hybrid cable may have a display issue and no image function is available
Root causeDevice Design
ActionSiemens issued a Urgent Medical Device Correction letter AX065/22/S on 3/21/23. Letter states reason for recall, health risk and action to take: has been performed. Please ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. Siemens will correct the hardware error via Update Instruction AX061/22/S. This measure will replace the current hybrid cable with a design improvement to enhance stability and quality of the signal transmission. Our service organization will contact you shortly to arrange a date to perform this correc
Quantity4 units US
DistributionWorldwide - US Nationwide distribution in the states of FL, SD, PR and the countries of Angola, Argentina, Australia, Brazil, Bulgaria, China, Georgia, Germany, India, Indonesia, Iraq, Japan, Malaysia, Morocco, Nepal, Nigeria, Poland, Russia, Rwanda, Serbia, Slovenia, South Africa, Spain, Sudan, S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133580ARTIS ONE
Siemens Medical Solutions USA, Inc., Apr 28, 2014