ClearanceBrief
Recall Z-1586-2023 ยท JAK System, X-Ray, Tomography, Computed

Philips North America

Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72813

DateMay 12, 2023 (initiated Mar 20, 2023)
StatusOpen, Classified
ReasonMulti-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury
Root causeUnder Investigation by firm
ActionPhilips Medical issued Urgent Medical Device Correction Letter Multi-Function Switch Unload Pedal Foot Entrapment on 3/20/23 to US consignees via Certified United States Postal Service (USPS) with delivery confirmation. International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Immediately stop using the Unload Pedal of the Foot Switch until Philips installs a solution on your system. " Instead of using the Unload Pedal, please use th
Quantity3 units US ;2 units OUS
DistributionWorldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, MI, MO, NY, TN, TX, VT and the countries of Australia, Chile, Costa Rica, Czech Republic, Dominican, Republic Germany, India, Israel, Japan, Lithuania, Philippines, Poland, Sint Maarten (Dutch part), South Africa, Thailand,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180015Philips Incisive CT
Philips Healthcare (Suzhou) Co., Ltd., Mar 20, 2018
K212441Philips Incisive CT
Philips Healthcare (Suzhou) Co., Ltd., Apr 27, 2022