ClearanceBrief
Recall Z-1551-2023 ยท KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer

Linkbio Corp.

LINK SLED Knee System POREX ENDO MODEL SLED TIBIAL COMP 13 MM X 55 MM Item Number: 15-2230/12

DateMay 4, 2023 (initiated Mar 21, 2023)
StatusOpen, Classified
ReasonAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
Root causeLabeling design
ActionLinkBio issued Urgent Medical Device Correction letter on March 21, 2023 to US Distributors. Letter states reason for recall, health risk and action to take: Please take note of the amendment / reinforcement of the Surgical Technique and participate in future offering of product training on the updated technique. Two warnings have been added. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to follow the updated surgical technique. Attendance in future product training on the LINK SL
Quantity5 units
DistributionNY, NV Foreign: Bulgaria, France, Italy, Norway, Poland, Sweden, Switzerland, Slovakia, Spain, United Kingdom, Germany, Austria, Argentina, Australia, China, India, Iraq, Jordan, Colombia, Mexico, Saudi Arabia, Ukraine, Vietnam, and Belarus
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152431LINK Endo-Model Knee System with PorEx (TiNbN) coating, Link Sled Knee System with PorEx (TiNbN) coating
Waldemar Link GmbH & Co. KG, May 9, 2016
K212742GEMINI SL Total Knee System, Endo-Model Knee System, and Sled Knee System with LINK PorEx (TiNbN) inhouse coating
Waldemar Link GmbH & Co. KG, Sep 16, 2021