ClearanceBrief
Recall Z-1428-2023 ยท JOH Tube Tracheostomy And Tube Cuff

Cook Incorporated

Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray, Reference Part Numbers C-PTISYJ-100-HC-G-EU-FLEX8.5 (G57725), C-PTISYJ-100-HC-G-NA-FLEX7.5 (G57720), C-PTISYJ-100-HC-G-NA-FLEX8.5 (G57721), C-PTISY-100-HC-G-NA-FLEX8.5 (G57717), C-PTISY-100-HC-G-NA-FLEX7.5 (G57716)

DateMay 10, 2023 (initiated Apr 7, 2023)
StatusOpen, Classified
ReasonThere were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.
Root causeNonconforming Material/Component
ActionAn URGENT: MEDICAL DEVICE RECALL dated 4/7/23 was sent to customers. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Sets and Trays from the Cook Medical lots in Attachment A. If so, quarantine any affected sets/trays that remain unused. Immediately cease all further distribution and use of the affected sets/trays. 2. Return the complete sets/trays to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form fo
Quantity25,161 units
DistributionWorldwide distribution - US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193133Blue Rhino G1-Multi Percutaneous Tracheostomy Introducer Set, Blue Rhino G1-Multi Percutaneous Tracheostomy Introducer T
Cook Incorporated, Dec 12, 2019