ClearanceBrief
Recall Z-1451-2023 ยท BZO Set, Tubing And Support, Ventilator (W Harness)

Draeger Medical, Inc.

VentStar Coax (P) 150, MP00379

DateMay 17, 2023 (initiated Apr 17, 2023)
StatusTerminated
ReasonGlued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Root causeProcess control
ActionThe firm distributed recall notifications to consignees via FedEx starting on 4/17/2023. Consignees are instructed to immediately stop use of affected devices, inspect current stock, and quarantine any affected devices to prevent use. Alternative therapy for patients should be instituted as soon as possible and ensure all users of affected devices within consignee organizations are aware of the recall, with copies being forwarded if devices were further distributed. To replace or return unused products, consignees are to call Draeger Customer Success between 8:00 AM - 6:00 PM EST, Monday thru
Quantity280 units
DistributionDomestic: US Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.