Recall Z-1645-2023 ยท DSJ Alarm, Blood-Pressure

Philips North America Llc

EarlyVue VS30 Vitals monitor with Software Versions A.00.02 and A.00.01 Product Number: 863380

DateMay 26, 2023 (initiated Mar 6, 2023)
StatusOpen, Classified
ReasonEarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.
Root causeSoftware design
ActionPhilips issued Urgent Field Safety Notice Letter on 3/3/23. Letter states reason for recall, health risk and action to take: 1. Identify the last calibration date of the device and review your hospital maintenance plan. Determine if your device has E01 End Tidal CO2 and has not been calibrated within one year of the production calibration. If needed adjust the maintenance plan and incorporate the calibration of your affected device. a. The device should not be used until the calibration or maintenance plan has been reviewed. b. The calibration date can be accessed via the following steps on th
Quantity1421 units US; 3 units OUS
DistributionNationwide Foreign: Belgium Brazil Canada Denmark Germany Guam Hong Kong Indonesia Italy Norway Poland Singapore Sweden Switzerland United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190624EarlyVue VS30
Philips Medizin Systeme Boeblingen GmbH, Oct 27, 2019