Recall Z-1395-2023 ยท HRS Plate, Fixation, Bone

GRAMERCY EXTREMITY ORTHOPEDICS

GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System

DateApr 17, 2023 (initiated Apr 4, 2023)
StatusOpen, Classified
ReasonInstrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.
Root causeMixed-up of materials/components
ActionOn 4/4/23, recall notices were mailed to customers and distributors who were asked to do the following: 1) Quarantine affected kits so they are ready for removal by the firm s sales reps or other contact. 2) If devices were transferred or further distributed, notify customers, or provide a customer list, so the recalling firm can notify these customers directly. 3) Complete and return the response form to msimpson@gramercyortho.com Customers with questions can contact the firm at 855-436-2278, 9AM 5PM, Monday - Friday.
Quantity24
DistributionUS Nationwide distribution in the states of CT, OR, NJ, AZ, CT, OH, AZ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200108GEO 1st MTP Joint Arthrodesis Plating System
Gramercy Extremity Orthopedics, LLC, Mar 26, 2020