Recall Z-1651-2023 ยท CBK Ventilator, Continuous, Facility Use

Philips Respironics, Inc.

Trilogy Evo O2 EV300, Model Numbers DS2200X11B, IN2200X15B, FX2200X15B, CA2200X12B, FR2200X14B, ND2200X15B, ES2200X15B, BL2200X15B, GB2200X15B, EU2200X15B, EE2200X15B, and TR2200X15B

DateJun 2, 2023 (initiated May 1, 2023)
StatusOpen, Classified
ReasonEnvironmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.
Root causeUnder Investigation by firm
ActionAn URGENT Medical Device Correction notification letter was sent to customers on 5/1/23. Please take the actions below in order to prevent risk for your patients. Actions that must be taken by the user in order to prevent risks for patients. To help prevent accumulation of debris on the machine flow sensor: Use the Philips approved particulate filter which prevents a significant majority of airborne aerosols and particulate from entering the device. This filter must be replaced between patients and monthly as indicated in the Instructions for Use. Using this filter was previously optional. Th
Quantity18,855 devices
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181166Trilogy Evo
Respironics, Inc., Jul 18, 2019