Recall Z-1498-2023 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM)- Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000

DateMay 2, 2023 (initiated Feb 2, 2023)
StatusOpen, Classified
ReasonIf, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range the C-Arm will reach the target position with an inaccuracy of 5-10 mm. As a result, the message Endposition reached will be displayed .Leading to unintended direction of the movement causing crushing of a patient, staff member, operator, or equipmen Live images may not match the previously stored reference images. Overlay images may be shown inaccurate on anatomy (e.g., DSA Road
Root causeSoftware design
ActionSiemens Healthineers issued to all users of ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM) a Customer Safety Advisory Notice to affected customers via AX041/22/S dated 2/3/23. Letter states reason for recall, health risk and action to take for addressing three potential software issues: Detection of wrong table movement direction; Adjust C-arm to Ref; Missing movement stop with multiple collision sensor activation. Siemens will correct the software error via Update Instructions AX040/22/S. Our service organization will contact you shortly to arrange a dat
Quantity141 units US
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov